All insights
Quality & Compliance3 min read

Traceability explained: the chain from the SI to your bench

Every calibration certificate claims traceability. Here is what that actually means, how the chain is built, and what breaks it.

Ask most engineers what makes a calibration valid and they will say it must be “traceable”. Ask them what traceability means and the answers become vaguer. It is worth understanding properly, because traceability is the property that makes a measurement in Ras Al Khaimah comparable with one made in Munich, Tokyo or Houston.

The formal definition

The International Vocabulary of Metrology (VIM) defines metrological traceability as the property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty.

Three ideas are packed into that sentence:

  1. There is a reference, ultimately the SI units.
  2. The chain is unbroken and documented: every link has a certificate.
  3. Each link has a stated uncertainty, which accumulates down the chain.

The chain in practice

For a typical DC voltage measurement, the chain looks something like this:

  • The SI volt is realised by national metrology institutes (NMIs) using Josephson voltage standards, which relate voltage to fundamental constants.
  • NMI calibration services calibrate high-level reference standards, such as Zener voltage references, for accredited laboratories.
  • Accredited laboratories calibrate the reference standards of labs like ours, including multifunction calibrators and reference multimeters.
  • Our reference standards calibrate our working standards, and our working standards calibrate your instruments.
  • Your instrument is then used to make measurements on your products or plant.

At each step the uncertainty grows slightly. A Josephson standard might realise the volt to a few parts in 10¹⁰; by the time we calibrate a handheld meter, the uncertainty might be a few parts in 10⁵. That is still far better than the meter’s specification, which is what matters.

What breaks the chain

Traceability is easy to lose. Common causes include:

  • Expired reference standards. If a lab’s standard is past its calibration due date, results produced with it are not traceable.
  • Missing uncertainties. A certificate that gives a measured value but no uncertainty does not support traceability, because the link’s contribution cannot be quantified.
  • Unaccredited or undocumented links. If a standard was calibrated by a supplier who cannot demonstrate their own traceability, the chain has an unknown link in it.
  • Undocumented corrections. Using an instrument with corrections from a certificate, but without recording which certificate, breaks the audit trail.

How to check

Your certificate should name the standards used and state their traceability. ISO/IEC 17025:2017 sets out how laboratories must establish and demonstrate traceability, and the ILAC policy on traceability (ILAC P10) describes acceptable routes to the SI. If in doubt, ask your provider to show you the chain for a specific parameter. A competent laboratory will be able to do that in a few minutes.

Start a calibration

Your next calibration is due sooner than you think.