A signal generator is a reference that your other measurements depend on. Here is how we verify that it produces what it says it does.

Signal generators are unusual among test instruments because they produce a quantity rather than measure one. Engineers use them as references to test receivers, characterise amplifiers and verify other instruments, so an error in the generator propagates silently into every measurement made with it.

Frequency accuracy

Like a spectrum analyzer, a signal generator’s frequency accuracy is set by its internal reference oscillator. We measure the output frequency, or the 10 MHz reference output, against our GNSS-disciplined standard. If the generator has an optional high-stability oven oscillator, we check it against that tighter specification.

Output level accuracy

This is usually the most demanding part of the calibration. Generators specify output level accuracy across both frequency and amplitude, often as a table with tighter limits at moderate levels and looser limits at very high and very low levels.

At high and moderate levels, we measure output power directly with a calibrated power sensor and meter. At low levels, down to −100 dBm and below, power sensors run out of sensitivity, so we use a combination of a calibrated step attenuator and a sensitive receiver or spectrum analyzer referenced back to the power-sensor measurement at a higher level.

Errors at low level matter. If you are testing receiver sensitivity at −110 dBm and the generator is actually producing −108 dBm, the receiver will appear 2 dB more sensitive than it really is.

Harmonics and spurious signals

A signal generator should produce a clean sine wave at the set frequency. We measure harmonics (at multiples of the carrier frequency) and non-harmonic spurious signals with a spectrum analyzer, and compare them with the specification, typically in dBc relative to the carrier.

Phase noise and residual FM

For generators used in communications, radar or local-oscillator applications, single-sideband phase noise at specified offsets is a key parameter. We also check residual FM and AM where specified.

Modulation

For generators with analog modulation, we verify the accuracy of AM depth, FM deviation and phase modulation at representative rates. For vector signal generators, we can verify modulation quality metrics such as error vector magnitude (EVM) on request.

Function and arbitrary waveform generators

Lower-frequency function and arbitrary waveform generators are checked for frequency, amplitude accuracy, DC offset, flatness across frequency, and the timing characteristics of square and pulse outputs. These are often used as references in production test and education, and they drift just like their RF cousins.

Look after the output connector

The most common physical fault we find on signal generators is a worn or damaged output connector. A precision N-type or 3.5 mm connector has a finite mating life and is easily damaged by over-tightening or by mating with a poor-quality cable. Use a torque wrench, keep connectors capped, and consider fitting a connector saver to instruments that are connected and disconnected many times a day.