RF power sensors and the calibration factor table
A power sensor's accuracy depends on a table of correction factors across frequency. If that table is wrong or out of date, every reading is wrong too.
RF power measurement sits at the base of almost every RF calibration chain. Signal generators, spectrum analyzers, attenuators and amplifiers are all verified, directly or indirectly, against power sensors. Getting power right is therefore not just about one instrument; it underpins a whole lab’s worth of measurements.
How a power sensor works
There are three main types of sensor in common use:
- Thermal sensors (thermocouple or thermistor) respond to the heating effect of RF power. They are accurate for any waveform shape but have limited dynamic range.
- Diode sensors rectify the RF signal. They are fast and sensitive, down to around −70 dBm, but are only true-RMS within their square-law region at low power.
- Peak and average sensors combine diode detectors with fast sampling electronics to measure modulated and pulsed signals.
In every case, the sensor’s response varies with frequency, because of mismatch and losses in the sensor element and its connector.
The calibration factor
To correct for this, each sensor is characterised across its frequency range and assigned a calibration factor (often written as Kb or CF) at each frequency. The calibration factor relates the power the sensor indicates to the power actually incident on it.
The power meter, or the sensor itself for modern “smart” sensors, applies the factor for the frequency you enter. If you forget to set the frequency, or the factor table is wrong, the reading can be in error by several tenths of a dB, and much more at high frequencies.
What a sensor calibration involves
A full sensor calibration determines:
- Calibration factors at a set of frequencies across the sensor’s range, typically by comparison with a reference sensor of known calibration factor, using a power splitter or a direct comparison system.
- Reflection coefficient (or VSWR) at each frequency, which is needed to calculate mismatch uncertainty when the sensor is used.
- Linearity across the sensor’s power range, comparing its response at different power levels against a reference.
The new factors are then either written into the sensor’s internal memory or supplied as a table for entry into the meter.
Calibrating the power meter
The power meter itself is calibrated separately. This checks the meter’s instrumentation accuracy, using a power meter calibrator that simulates sensor outputs, and its 50 MHz reference output, which is used to zero and reference the sensor before measurements. The 50 MHz reference is typically 1 mW (0 dBm) with a tight specification, and an error here offsets every measurement you make.
Mismatch: the hidden uncertainty
When a source and a sensor are both imperfectly matched to 50 Ω, some power is reflected back and forth between them. This creates mismatch uncertainty, which can easily be the largest term in a power measurement uncertainty budget. It is calculated from the reflection coefficients of the source and the sensor, which is why sensor reflection data is reported on our certificates.
Looking after sensors
Power sensors are easily damaged by overload, and their connectors wear. Always check the maximum input power, use an attenuator on high-power sources, and use a torque wrench when connecting. A damaged sensor connector can also damage every port it is mated with.