All insights
Electrical3 min read

Clamp meters and insulation testers: a guide for electrical contractors

The two instruments every electrical contractor relies on, and the two most likely to come back from site damaged. What calibration covers and how to keep them reliable.

Electrical contractors and maintenance teams make up a large share of our customers, and two instruments dominate their kit: the clamp meter and the insulation resistance tester. Both are used to make decisions about safety, and both lead hard lives. Here is what their calibration involves.

Clamp meters

A clamp meter measures current by sensing the magnetic field around a conductor, without breaking the circuit. AC clamps use a current transformer; AC/DC clamps use a Hall-effect sensor.

What we calibrate

  • AC and DC current, using a calibrator and a multi-turn current coil. The coil multiplies the calibrator’s output: a 50-turn coil carrying 20 A presents 1,000 A to the clamp.
  • Jaw position sensitivity, by checking readings with the conductor at different positions in the jaw. Worn or misaligned jaws cause errors that vary with position.
  • Voltage, resistance and continuity functions, which are calibrated as for a multimeter.
  • Frequency response, where the clamp is specified for harmonics or true-RMS measurement of distorted waveforms.

Common faults

Dirt or corrosion on the jaw faces is the most common cause of error on AC clamps, because it opens a gap in the magnetic circuit. Hall-effect clamps are also sensitive to residual magnetism and need to be zeroed before DC measurements. If a clamp reads differently each time you open and close it, check the jaw faces.

Insulation resistance testers

Insulation testers apply a high DC voltage, typically 250 V to 5 kV, and measure the resulting leakage current to calculate insulation resistance, often into the gigaohms.

What we calibrate

  • Test voltage, measured with a high-voltage divider or meter, at each selectable range. The open-circuit voltage should be within specification, and should be maintained under load. IEC 61557-2 sets requirements for the test voltage under a rated load.
  • Resistance measurement, using high-value standard resistors or a dedicated insulation tester calibrator, across the range from megohms to hundreds of gigaohms.
  • Short-circuit current, which determines how quickly the tester can charge capacitive loads such as long cables.
  • Polarisation index and dielectric absorption ratio timing, on testers that offer these functions.
  • Continuity and low-resistance functions, where fitted.

Common faults

High-value resistance measurements are very sensitive to surface leakage. Dirty or damp test leads can produce falsely low readings, which may cause good insulation to be rejected. Lead damage is common and is a safety issue at these voltages: inspect leads before each use and replace any with damaged insulation.

How often?

For both instruments, twelve months is the norm. If your meters are used daily on construction sites or in industrial plants, consider a check against a simple reference, such as a known resistor box for insulation testers, at the start of each job. It takes a minute and can prevent a costly mistake.

Start a calibration

Your next calibration is due sooner than you think.

Send us your instrument list and we will come back within one working day with a fixed quote, turnaround date and collection slot.