As-found vs as-left: why you should insist on both
The as-left result tells you the instrument is good now. The as-found result tells you whether everything you measured last year was good. Only one of them is about the past.
If you take one thing from this blog, let it be this: always ask for as-found data. It is the single most useful piece of information a calibration can give you, and it is the first thing lost when a lab adjusts an instrument without recording it.
The two snapshots
As-found data is the instrument’s performance as it arrived at the lab, before any cleaning, repair or adjustment.
As-left data is its performance after any adjustment, as it leaves the lab and goes back into service.
If no adjustment is made, the two are the same, and a single set of results is reported.
Why as-left is not enough
An as-left result tells you the instrument is in tolerance today. That is useful, but it says nothing about the last twelve months.
Imagine a power meter that has drifted 0.4 dB out of tolerance over its calibration interval. If the lab adjusts it back into specification and reports only as-left data, the certificate shows a perfectly good instrument. Meanwhile, a year’s worth of transmitter power measurements, acceptance tests and customer reports may have been wrong, and nobody knows.
What as-found data lets you do
With as-found data, you can:
- Assess the impact of a failure. If an instrument was out of tolerance, you can identify which measurements it was used for and decide whether they need to be reviewed or repeated.
- See drift. Comparing as-found results across several calibrations shows whether an instrument is stable or drifting, and in which direction.
- Justify interval changes. Stable as-found history supports extending an interval; repeated failures argue for shortening it.
- Satisfy auditors. ISO 9001:2015 clause 7.1.5.2 requires organisations to determine whether the validity of previous results has been adversely affected when equipment is found unfit for purpose. You cannot do that without as-found data.
How we handle it
At PME Laboratory, as-found data is always recorded, whether or not you ask for it. If an instrument is found out of tolerance, we:
- Notify you the same day with the as-found results.
- Ask whether you want the instrument adjusted, repaired or returned as found.
- If adjusted, perform a full calibration afterwards and report as-left results alongside the as-found data on the same certificate.
A note on “limited” calibrations
Some providers offer cheaper calibrations that skip as-found measurements and go straight to adjustment. It saves time on the bench, but it removes the most valuable part of the service. If a quote looks unusually low, ask whether as-found data is included.